7. Cleaning and Maintenance
Step
Daily
Weekly
Monthly
every
3 months
Check the filter screens of the paraffin
reservoirs and clean if necessary.
√
Check the air circulation opening of the
paraffin reservoirs and clean if necessary.
√
Check and empty the condensate container.
Clean the inlet openings.
√
Check the status of the active carbon filter
.
√
Replace the active carbon filter.
√
Check the level sensors and clean if
necessary.
√
Check the electrical ports on the rear
side of the instrument.
√
Clean the reagent containers from the
inside if necessary.
√
Make sure that the shutter caps of the
reagent containers and the O-rings fit
correctly and are tight.
√
Check that the reagent containers are
fitted securely in the ports.
√
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)