7.2.2 Periodic cleaning and maintenance (continued)
Cleaning the level sensors
44
Fig. 86
Fig. 87
Use the cleaning tool together with the microfiber cloth (
Fig. 85
) as shown in
Fig. 87
.
45
46
47
7. Cleaning and Maintenance
The level sensors must be cleaned of all residues (reagents and paraffin) after every
cleaning program!
NEVER use the cleaning tool WITHOUT the accompanying microfiber cloth, as other-
wise the sensors will be scratched!
The cleaning tool has two different spoons. The larger spoon (
Fig. 85
, 46) is for use
with the ASP6025, the smaller (
Fig. 85
, 47) is for use with the ASP200 S / ASP300 S.
To prevent damage to the sensors, NEVER use the larger spoon for cleaning the
sensors in the ASP200 S / ASP300 S.
You will be prompted to clean the sensors by the instrument software at the end of each
cleaning program (
Fig. 86
, 44).
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)