5.1.5 Viewing the program list
You can create:
• Up to 15 infiltration programs
• 3 retort cleaning programs
• 1 paraffin cleaning program
At supervisor access level:
• The program names for the infil-
tration programs can be edited.
• New infiltration programs can be
added and existing ones can be
deleted.
Fig. 43
→
This list (
Fig. 43
) shows all programs defined in the ASP300 S.
5. Operation
Important!
New infiltration programs are created by copying an existing pro-
gram. Therefore, the list must contain at least one program at all
times.
Retort or paraffin cleaning programs are preset. They cannot be
renamed, added or deleted.
Program duration cannot be specified. It is determined by the total duration
of all program steps, plus the estimated fill and drain times. To alter the
duration of a program, the duration of one or more individual program steps
has to be modified.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)