Cleaning and maintenance – Leica Biosystems ASP300 S User Manual
Page 77
78
Leica ASP300 S
Replacing the active carbon filter
• The life of the active carbon filter will depend
on the reagent types used and the frequency
of vacuum cycles.
• The filter (13) should be replaced at least every
3 months (
Fig. 84
).
After replacing the filter, open the SMART
MONITOR menu to reset the filter age to "0".
For additional information, refer to
Chap. 5.5
• We recommend a smart clean prior to chang-
ing the filter.
Fig. 84
Draining the condensate container
• Remove, empty, and reinsert the blue conden-
sate container (
Fig. 83
).
Fig. 83
Filter label
7. Cleaning and Maintenance
14
13
Dispose of waste solvents with care
according to local regulations and the
waste management policy of the com-
pany or institution.
When inserting the new filter, make
sure that it is inserted with the correct
side up, as specified on the label on
the front side of the filter. The arrow
must point upwards.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)