3 running programs, Running programs, Operation – Leica Biosystems ASP300 S User Manual
Page 60
61
Instructions for Use, V 1.5 RevD – 03/2013
5. Operation
5.3 Running programs
There are two different windows from which a program can be started:
In the Start screen, press the ALL
PROGRAMS button.
This window displays all of the
infiltration programs defined in the
instrument.
Up to five favorites can be configured in the
ASP300 S.
A "Favorite" is an infiltration program that is used
frequently and thus has been assigned to the
FAVORITES list.
All settings have already been programmed and
only the number of blocks needs to be entered (if
this function has been activated).
Starting a "Favorite"
To begin working with a Favorite, press the
corresponding symbol on the start screen. The
program is started immediately.
After starting, the end time or other program op-
tions can be changed in exactly the same way as
for any other running program.
FAVORITES
ALL PROGRAMS
Fig. 61
Fig. 62
• To call up a program, touch the button that is
labeled with the program name.
• A screen with the graphic representation of
the program appears; in it, all of the stations
necessary for this program are color coded
according to reagent group (
Fig. 63
).
• Unlike a Favorite, here you have the ability to
modify a program before it is started.
However, all modified processing steps apply
only for the program that is currently called up.