Leica ASP300 S
5.2.2 Replacing the paraffin
Draining the paraffin
• Connect the paraffin drain hose (36) to the
spout (17) of the paraffin drain at the front of
the instrument (
Fig. 55
) and into a catch tank.
Fig. 55
• Select the paraffin reservoir to be emptied and
the receptacle (External,
Fig. 56
).
Then, press STATION DRAIN....
• This is followed by multiple security prompts
that you have to confirm individually in order
to continue.
17
36
From FAVORITES, call up the SMART SCREEN
screen. (see previous chapter)
→
+
→
Fig. 56
5. Operation
When connecting the hose, be sure to
press it onto the O-rings of the drain
opening as far as it will go.
The paraffin drain hose must be
securely hooked into the external
receptacle and remain there through
the entire drainage process.
When drainage is finished, the hose is
cleaned automatically using air.
Do not remove the hose from the exter-
nal receptacle until this cleaning step
is complete.