3 installing accessories, Installing accessories, Initial operation – Leica Biosystems ASP300 S User Manual
Page 26
27
Instructions for Use, V 1.5 RevD – 03/2013
4. Initial Operation
Fig. 5
• Move the instrument to its final setup location.
Active carbon filter
• Unpack the active carbon filter (13) and insert
it (see
Fig. 5
).
When doing so, make sure that the filter is
inserted with the correct side up. Note the
direction of the
label (29) on the
front of the filter –
the arrow must be
pointing upwards.
13
4.3 Installing accessories
18
29
Remote fill / drain hose
• Connect the supplied remote fill/drain hose to
the remote drain connection on the front of the
instrument (
refer Fig. 6
).
• Important!
When inserting the hose into the drain opening
(
Fig. 5/6,
18), the connecting device (28) of the
hose must engage with a clearly audible click.
Fig. 6
28
18
If it is possible to set up the bulk con-
tainer for filling and draining in the
immediate vicinity of the instrument,
the hose can be shortened so that it is
easier to handle.
When shortening the hose, you can cut
a V-shaped notch into the hose end to
attain better flow.