2 connecting an uninterruptible power supply (ups), Initial operation – Leica Biosystems ASP300 S User Manual
Page 25
![background image](/manuals/565440/25/background.png)
26
Leica ASP300 S
• Connect the ASP to a grounded socket using
the power cable (37).
• Using the jumper cable (38), connect the output
of the main voltage supply to the UPS input.
• Using the jumper cable (38) connect the input
of the electronics section (40) to one of the
UPS outputs.
• Switch the instrument on at the main switch.
• Start the UPS.
4.2.2 Connecting an uninterruptible power supply (UPS)
38
40
37
38a
Fig. 4a is provided as an example only
and shows the correct connection
of the ASP300 S to an uninterruptible
power supply (UPS).
Fig. 4a
An uninterruptible power supply (UPS)
protects machines and instruments
against malfunctions in the power
supply.
Leica recommends using an active
tracking UPS (with an output power of
1000-1500 VA) to protect the instrument
and the specimens from temporary
power failure, voltage spikes, under-
voltages and excess voltages.
4. Initial Operation