1 editing a program that has been called up, Operation – Leica Biosystems ASP300 S User Manual
Page 61
62
Leica ASP300 S
5.3.1 Editing a program that has been called up.
Fig. 63
To modify a program before the start, press
the EDIT button. A message appears that the
changes will affect running programs only. After
confirming with YES, the program edit window
appears (
Fig. 64
).
Fig. 64
Resetting the end time:
• Touch the FINISH TIME button and reset the
end time in the input window.
Fig. 65
↓
• Enter the day on which you want the program
to end.
• Enter the time (in 24-hour format) at which you
want the program to end.
• Press OK to confirm.
→
↓
5. Operation
Programs can be started up to 6 days
in advance. So that the program is
ended at the defined date/time, the
delay step is lengthened accordingly.
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)