5.4 Reagent status
This window (
Fig. 64
) is used to display and update the current reagent
status of the stations. In addition, reagent stations can be marked as
full or empty here.
→
Fig. 75
The following is displayed:
• The current age of each reagent
• The current status of each station
• Criteria that trigger warning mes-
sages (Fields colored in red)
To change the status of a reagent,
highlight the corresponding row
and press the button symbol for the
respective action.
In addition, you can switch directly
to the REAGENTS and STATIONS
windows to make current changes.
Reagent warning messages
If the RMS is active, warning messages are out-
put if reagents are used for too long.
The output of the warning messages can take
place after up to 3 criteria for normal reagents
and after up to 6 criteria for paraffins.
Warning messages are displayed at the end of a
cleaning program. Then, you can switch directly
to the REAGENT STATUS display.
Warning messages also appear (with a yellow
background) in the windows in which programs
are started.
The criteria for warning messages
For normal reagents, it is the number of the
following since the last reagent change:
• Edited blocks
• Run programs
• Days passed
For paraffins, warning messages can also be
output according to the number of the following
since the last paraffin cleaning:
• Edited blocks
• Run programs
• Days passed