Operation – Leica Biosystems ASP300 S User Manual
Page 37
38
Leica ASP300 S
PROGRAM OPTIONS
Select the options with which programs are executed here. Press a button
to change the associated value.
• STATION SEQUENCE:
By Age or Sequential
By Age -
RMS is activated, reagents will automatically be
used in order of increasing cleanliness.
Sequential -
reagents will be used in sequential order of stations.
• WAX BATH ORDER:
Auto or 1. ; 2. ; 3.
Can only be activated if STATION SEQUENCE is set to "Sequential" –
otherwise the order of baths will be managed by the RMS.
• PROMPT FOR NUM BLOCKS: Enabled or Disabled
Enabled -
When starting a program, the actual number of
blocks must be entered. This is recommended with
the RMS being enabled.
Disabled -
Number can be entered, but does not have to be.
DATE / TIME
Be sure to verify that the date and time entered do actually correspond
to the local time/date, as this will ensure that all programs are carried out
correctly.
Press the buttons to access the entry windows where settings or correc-
tions can be made, if necessary.
Date and time setup screen
Fig. 26
Fig. 24
Fig. 25
5. Operation