50
3.6 CompactFlash card
Automatic
storage of
measurement
data
The data stored in the measurement data memory (FLASH) of the paperless
recorder are saved at regular intervals to the CompactFlash card
1
in the
instrument. The PC evaluation program (
v Chapter 7 “PC programs”) reads
the data from the CF card and provides user-friendly functions for evaluation.
Loading and
saving the
configuration
data
The configuration data can be downloaded from the CF card and saved to the
CF card. A configuration can thus be copied from one instrument to another,
or it can be transferred from and to the PC setup program.
Basic menu ➔
CompactFlash
card
The CompactFlash menu is called up via the basic menu:
The functions
- Update CF incl. counters ...,
- Compl. meas. data ➔ CF,
- Config. data ➔ CF card and
- CF card ➔ Config. data
are protected against unauthorized access by a password query (factory
setting: 09200).
1
PC Card access made available by CSM FAT File System
Copyright © 1997-2002 CSM GmbH Filderstadt, Germany
A
Only one configuration file can be saved to the CF card at any
time. If there is a configuration on a CF card, this will be
overwritten without any further query through the function
CompactFlash card ➔ Config. data➔CF card.
h Select operating level CompactFlash card
h Enter selection with E