2 Instrument Description
If the measurements go outside the tolerance band, and the new
measurement, which is outside the tolerance band, is present for at least the
duration of the storage cycle that was set in the active operating mode, then it
is acquired and stored and a new tolerance band is applied.
H
Going outside the tolerance band is not necessarily recorded. If
the measurements return inside the tolerance band within the
storage cycle (t2), then there will be no recording and “Min. storage
cycle” is still active.
10s
10s
10s
10s
10s
10s
10s
10s
t
t1
t2
t3
t4
10s
t6
t5
t1 = storage through “Min. storage cycle” (forced storage) and
application of a new tolerance band.
t2 = no storage, since the measurement is again within the tolerance
at the end of the storage cycle (10 sec).
t3 = storage and application of a new tolerance band, since the
measurement has gone outside the tolerance band at the end of
the storage cycle.
t4 = storage and application of a new tolerance band, since the
measurement has gone outside the tolerance band at the end of
the storage cycle.
t5 = storage through “Min. storage cycle” (forced storage) and
application of a new tolerance band.
t6 = no storage, since the measurement is again within the tolerance
at the end of the storage cycle (10 sec).