21 configuration - counters/integrators – JUMO 706585 LOGOSCREEN fd Operating Manual User Manual
Page 156

21 Configuration - Counters/Integrators
156
With the function Flow rate quantity, the time base determines the counting
duration of pulses (gate time) and display interval.
Threshold
Threshold value is only available for the Function “Integrator”. An integration
can only be made if the present measurement is larger than the threshold
value. Time base and weighting are not considered in the threshold value
comparison.
Reset
signal
In addition to the normal end, a signal can be specified here that can be used
to set the selected counter and integrator count to 0.
This parameter is not available for the function Flow rate quantity.
Channel name
(designation)
Short designation for the counter/integrator, up to 7 characters. The channel
name is displayed in the individual visualization modes, together with the
channel designation.
Channel
description
Description of the counter/integrator, up to 21 characters. The channel
description is displayed in the individual visualization modes, together with the
channel name.
Unit
Unit of measure in which the counter or integrator count will be displayed, up
to 5 characters. The unit is displayed wherever the count is represented
numerically.
Decimal place
(Comma
format)
The decimal place serves to determine the number of integer and decimal
places for the numerical representation of counter states. If required, the
instrument will automatically switch to a different format, in order to be able to
show all the integer places. As a rule: all integer places must be displayed.
Second
The flow rate quantity is determined over the course of a
second and displayed.
Minute
The flow rate quantity is determined over the course of a
minute and displayed.
Hour
The flow rate quantity is determined over the course of a
hour and displayed.
Day
The flow rate quantity is determined over the course of a
day and displayed.
H
The current status is not saved. The reset is performed in addition
to the Reset signal from Chapter 21.1 "General Setting".