MDC-360C DEPOSITION CONTROLLER
COMPUTER INTERFACE
11-14
Example: To instruct the controller to send the source setup parameter list the
computer would send:
Chr$(255)+Chr$(254)+Chr$(1)+Chr$(18)+Chr$(0)+Chr$(237)
20. Receive source setup (Code #19)
Instructs the controller to enter the incoming source setup parameters into
memory. A description of the source setup parameter list is given above.
21. Send sensor setup (Code #20)
Instructs the controller to send the sensor setup parameter list for all sensors. A
description of the sensor parameter list is as follows:
Parameter Name
Length
(bytes)
Decimal Pt.
Position
Range
Units
Number of crystals
1
*
(1-8)
None
Shutter Relay Type
1
*
44=Normally open,
45=Normally closed,
46=None,
47=Dual
None
Control 1
*
68=Direct,
69=BCD,
70=Indiv,
71=Manual
None
Drive 1
*
60=Up,
61=Down,
62=Fast,
63=Inline,
64=Sngl Step,
65=Dbl Step
None
Feedback input type
1
*
8 =Individual,
9=BCD,
10=Single Home,
11=In Position,
12=No Feedback
Rotator Delay
1
*
(0-99)
Seconds
Total 6 bytes
There are a total of six sensors with 6 bytes per sensor for a total of 36 bytes. All
36 bytes are sent in the same message starting with sensor #1.
Example: To instruct the controller to send the sensor setup parameter list the
computer would send:
Chr$(255)+Chr$(254)+Chr$(1)+Chr$(20)+Chr$(0)+Chr$(235)
22. Receive sensor setup (Code #21)