Instruction descriptions, Instruction, Descriptions -5 – INFICON MDC-360C Thin Film Deposition Controller User Manual
Page 147

MDC-360C DEPOSITION CONTROLLER
COMPUTER INTERFACE 11-5
36
Send Process Log Directory
37
Send Process Log Data
38
Initiate Remote Action Command
39
Send Individual Run-time Value Command
40
Enable/Disable the front panel keyboard
11.9 INSTRUCTION
DESCRIPTIONS
The following is a description of all the valid instructions along with an example
of how they are used. All the examples assume the controller address is 1.
1. Remote activation of controller (Code #0)
This instruction initiates a key press of the MDC-360C’s keyboard. The valid key
codes are shown in the following table:
Remote Activation Code
Description
1 Program
key
2 Manual
key
4 Shutter
key
8 Zero
key
16 Reset
key
32 Abort
key
64 Start
key
Format: Header, Instruction=1, Length=1, Key Code, Checksum
Example: To initiate a zero thickness instruction the computer would send:
Chr$(255)+Chr$(254)+Chr$(1)+Chr$(0)+Chr$(1)+Chr$(8)+Chr$(246)
2. Send controller hardware configuration (Code #1)
Instructs the controller to send controller configuration data to the host computer.
The following is a description of the configuration data:
Name
Length (bytes)
Message
Software Version
30
MDC-360C Software Version X.X
Source/Sensors Cards
1
(1, 2 or 3)
I/O Ports
1
(1 or 2)
Communication Port
1
(1=RS232, 2=RS-485, 3=IEEE488)
Data Log Storage
1
(0=No, 1=Yes)
Total 34 bytes
Example: To instruct the controller to send the hardware configuration data the
computer would send:
Chr$(255)+Chr$(254)+Chr$(1)+Chr$(1)+Chr$(0)+Chr$(254)
3. Send utility parameters (Code #2)