beautypg.com

Chapter 5 troubleshooting, 1 troubleshooting tools, 1 symptom, cause, remedy – INFICON Front Load Single Sensor User Manual

Page 77: Chapter 5

background image

5 - 1

PN

07

4-

15

6N

Front Load Single and Dual Sensors Operating Manual

Chapter 5

Troubleshooting

5.1 Troubleshooting Tools

If the Front Load Sensor fails to function, or appears to have diminished
performance, diagnose the sensor using one or more of the following:

Symptom, Cause, Remedy chart (see

section 5.1.1

)

Digital Multimeter (see

section 5.1.2 on page 5-3

)

Crystal Sensor Emulator (see

section 5.1.3 on page 5-10

)

5.1.1 Symptom, Cause, Remedy

The Symptom, Cause, Remedy chart can help identify the causes of, and solutions
to, sensor problems and related issues (see

Table 5-1

).

Table 5-1 Symptom, Cause, Remedy

SYMPTOM

CAUSE

REMEDY

Large jumps of thickness
reading during deposition.

Mode hopping due to damaged
or heavily damped crystal.

Replace the crystal.

Crystal is near the end of its life.

Scratches or foreign particles on
the crystal holder seating
surface.

Clean or polish the crystal
seating surface of the crystal
holder. Refer to

section 4.3.2 on

page 4-6

.

Uneven coating.

Mount the sensor with the
crystal face perpendicular to the
evaporant stream. Refer to

section 2.2 on page 2-4

.

Particles on the crystal.

Remove source of particles and
replace the crystal.

This manual is related to the following products: