INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 97
![background image](/manuals/562454/97/background.png)
Protec P3000 Messages
5-7
k
in
a
26
e Ch
ap
te
r
5
.fm
Tech
ni
ca
l Ha
nd
bo
ok
(1
20
9)
W65
Wrong date set!
Wrong date in the Protec.
Wrong Hex code entered to
PRO-Check
Check the date in the Protec.
Check the Hex code entered to
PRO-Check
W66
New PRO-Check!
New PRO-Check reference leak
installed in main unit
Please enter serial number and code.
See Section
, Replace PRO-Check
W67
PRO-Check will be expired on
DD.MM.YYYY
Warning for pending expiration will
be issued adjusting 14, 30, 60, or
90 days before actual expiration.
Please order new PRO-Check gas
reservoir! (cat. no. 521-010)
See Section 7.7.5
W68
PRO-Check expired!
PRO-Check has been used for
more than 1 year or has been
manufactured more than 2 years
ago.
Please replace PRO-Check gas
reservoir!
See Section
W70
All EEPROM parameter of
PRO-Check lost!
EEPROM in PRO-Check is empty
or faulty
Replace PRO-Check!
W71
No communication with PRO-
Check!
No or defective electrical
connection between PRO-Check
and main unit
Check connection of PRO-Check with
main unit
If problem persists, call nearest
INFICON service repesentative!
PRO-Check not installed in main
unit
Install PRO-Check!
No PRO-Check available
Disable PRO-Check in software menu
(see Section
)
W72
No communication with sniffer!
No or defective electrical
connection between sniffer line
and main unit
Check connection of sniffer line with
main unit (disconnect and reconnect, if
possible, try different sniffer line)
If problem persists, call nearest
INFICON service
W78
Signal difference between
calibrated leak and air too
small!
Calibrated leak too small / empty
during calibration or proof
Check leak rate of test leak or use leak
with higher leak rate
Background too high during
calibration
Check helium background
See Section
, Info page 2.
Not enough time expired for air
signal to stabilized (acknowledged
too early)
Repeat calibration, allow sufficient time
for air signal to stabilize!
W81
Calibration factor too low!
Calibration factor determined to be
< 0.1 during calibration
Leak rate of test leak is incorrect
(especially during external
calibration)
Check correct setting of test leak value!
W82
Calibration factor too high!
Calibration factor during calibration
determined to be > 10
Leak rate of test leak is incorrect
(especially during external
calibration)
Check correct setting of test leak value!
Test leak not sniffed properly / not
long enough
Repeat calibration, sniff test leak
properly and for sufficient time
Error
no.
Plain text message
Possible Reason
Recommendation