Sniffer tip filter
The maintenance interval for the sniffer tip filter depends on the operating
environment and may be set to values between 10 and 999 hours. If set to
∞
the
reminder message for the sniffer tip filter will be disabled.
Default value: 100 hours
For details on how to perform the maintenance tasks see Chapter
, Maintenance.
When confirming that the filter has been replaced (without changing the interval time)
press okay only and the counter will restart.
Maintenance intervall
In the maintenance interval sub-menu the number of operating hours for the main
unit in total, as well as the time until the next maintenance is due for the diaphram
pump, the main unit air filter and the optimization for the Wise Technology sensor.
When the number of operating hours have expired a warning will be issued as a
reminder to perform the appropriate task. The reminders are issued for the
diaphragm pump and the air filter of the main unit. Please also refer to Section
(Error Messages and Warnings)!
Fig. 4-22 Setting the maintenance interval for the sniffer tip filter and confirming its
replacement
Fig. 4-23 Example of a maintenance interval menu page