Converting a trigger leak
rate to a trigger output
voltage

If you want to know the corresponding voltage for a specific trigger leak rate, the
following formula applies:

U = 3 + 2·log (mantissa of scientific notation of trigger value)

(The factor of 2 is used since one decade is spread over 2 Volts. The summand of 3
is used because the trigger level will always be within the range of 3 to 5 V).

Example

Recorder Output

In linear output scale different settings can be chosen in the submenu

RECORDER

OUTPUT

. The voltage range spans from 0 V to 10 V.

The recorder output can be set so that the max. 10 V corresponds to either 1 x 10

-4

mbar l/s, 1 x 10

-3

mbar l/s, 1 x 10

-2

mbar l/s, 1 x 10

-4

Torr l/s, 1 x 10

-3

Torr l/s,

1 x 10

-2

Torr l/s. Alternatively, the recorder otuput can be set so that 10 V

corresponds to 10 x trigger level, i. e. 1 V equals the selected trigger level.

In the modes "E

RROR

”, "N

OT

READY

TO

MEASURE

” and "

STANDBY

” a voltage of

U = 10 V will be output through channels 1 and 2.

Default setting:

AUTO

4.5.3

RS232 Protocol

The RS232 protocol can be set to “

ASCII

”, “

DIAGNOSTICS

”, “

PRINTER

AUTO

” or “

PRINTER

MANUAL

”. The ASCII protocol is a protocol similar to the SCPI, a standard protocol

widely used for measuring equipment. For details on this protocol please refer to the
Interface Description (kins26e1).

Default setting: ASCII

Diagnostics

The

DIAGNOSTICS

protocol is a binary protocol used by INFICON for analysis in quality

control and trouble shooting during manufacturing and service.

Trigger leak rate

Scientific notation

Mantissa

Trigger voltage

2*10

-5

mbarl/s

2*10

-5

2

3+2·log(2) = 3.6 V

5 g/a

5*10

0

5

3+2·log(5) = 4.4 V

0.4 oz/yr

4*10

-1

4

3+2·log(-1) = 4.2 V