Setting-up / editing an i•guide program, 4 setting-up / editing an i•guide program – INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 74

4-19
Equipment Settings
k
in
a
26
e Ch
ap
te
r
4
.fm
Tech
ni
ca
l Ha
nd
bo
ok
(1
20
9)
4.4.4
Setting-up / editing an I•Guide Program
When selecting the
SETUP
PROGRAM
button a submenu page will open showing the
list of 10 pre-programmable I•Guide programs and their current status (enabled /
disabled). If a program is enabled it will be offered in the list of programs to select
from when operating in the I•Guide mode. To change the status of a program use the
ON
/
OFF
button on the left side of the display.
To edit the parameters of a program select the program by scrolling to the
corresponding menu line and open the selected program by pressing the “E
DIT
”
button.
In the opening
EDIT
PROGRAM
sub-menu the name of the program, the gas type
selected for this program, the number of locations to be leak tested, the minimum
measuring time for each location as well as the wait time between two leak tests
allowed for moving the sniffer to the next location is stated. In addition, the
summarized global trigger value corresponding to the allowable leak rate for the
complete unit under test is stored.
Each parameter can be edited by scrolling to the appropriate line item and selecting
the highlighted item by pressing the
EDIT
button.
Fig. 4-17 Setting-up the I•Guide Mode
Fig. 4-18 Editing an I•Guide program