Protec P3000 Messages

5-5

k

in

a

26

e Ch

ap

te

r

5

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

W42

Pressure at calibration differs
between

HIGH

and

LOW

FLOW

For Protec P3000XL only in

HIGH

FLOW

mode:

Pressure at Wise Technology
sensor differs from

LOW

FLOW

mode to high flow mode (during
start-up or calibration only)

Filter in sniffer line blocked

Internal filter in main unit blocked

Capillary blocked

Replace sniffer tip filter and recalibrate
(See Section

7.4

)

Replace internal filter* and recalibrate

Replace sniffer line and recalibrate!
Or: Replace probe cable!* and
recalibrate

See Section

7

W53

Flow at calibration out of limits!

Only for Protec P3000 or Protec
P3000XL in

LOW

FLOW

mode:

Flow at calibration higher than
upper flow limit or lower than lower
flow limit or limit low flow error (if
enabled, whichever comes first)

Filter in sniffer line blocked!

Internal filter in main unit blocked

Leak in (low flow) capillary

Wrong flow limit or limit low flow
error settings (for

LOW

FLOW

mode)

Replace sniffer tip filter and recalibrate!
(see section 7.4)

Replace internal filter* and recalibrate!

Replace sniffer line (or: replace probe
cable*) and recalibrate!

Check flow limit settings and limit low
flow error (for

LOW

FLOW

mode)! See

section

4.4.1

W54

Flow at calibration out of limits!

Only for Protec P3000XL in

HIGH

FLOW

mode:

Flow at calibration higher than
upper flow limit
or lower than lower flow limit or limit
low flow error (if enabled,
whichever comes first)

Filter in sniffer line blocked

Internal filter in main unit blocked

Leak in probe cable

Wrong flow limit or limit low flow
error settings (for

HIGH

FLOW

mode)

Replace sniffer tip filter and recalibrate!
(see section 7.4)

Replace internal filter* and recalibrate!

Replace sniffer line (or: replace probe
cable*) and recalibrate!

Check flow limit settings and limit low
flow error (for

HIGH

FLOW

mode)! See

section

4.4.1

Error
no.

Plain text message

Possible Reason

Recommendation