Maintenance

7-12

k

in

a

26

e Ch

ap

te

r

7

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

The steel capillary may be removed in that way for the following sniffer tips:

3

Take out the steel capillary and remove the taper gasket at the top of the sniffer tip.

Notice

The steel capillary can be cleaned with compressed air or a thin steel wire.

4

Reinsert the steel capillary and reinstall the sniffer tip.

5

Screw the plastic capillary filter onto the sniffer tip.

6

Please recalibrate the Protec P3000 for most accurate display of leak rate.

7.6.2

Switching from plastic to metal capillary filter

Notice

When you switch from the plastic capillary filter to the metal capillary filter,
do not forget to re-install the taper gasket as otherwise the SL3000 sniffer
line will be leaky.

1

Unscrew the plastic capillary filter.

2

Insert taper gasket (Fig.

Fig. 7-14

/1)

3

Re-install the capillary in the sniffer tip.

4

Screw in metall capillary filter at the end of the sniffer tip.

5

Please recalibrate the Protec P3000 for most accurate display of leak rate.

Fig. 7-16 Steel capillary sticking out of sniffer tip flange

cat. no.

length

122 09

FT600

600 mm

flexible

122 13

ST312

120 mm

rigid

122 14

FT312

120 mm

flexible

122 15

ST385

385 mm

rigid

122 16

FT385

385 mm

flexible

122 18

FT200

200 mm

rigid

122 66

FT250

250 mm

flexible

122 72

ST500

500 mm

45° angled