Available configurations, Technical data, 2 available configurations – INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 12: 3 technical data
![background image](/manuals/562454/12/background.png)
1-2
General Information
k
in
a
26
e Ch
ap
te
r
1
.fm
Tech
ni
ca
l Ha
nd
bo
ook
(1
20
9)
1.1.2
Available Configurations
The Protec P3000 leak detector is available in four different configurations:
The Standard Protec
P3000
The standard Protec P3000 should be used for high sensitivity applications. It
requires the SL3000 sniffer line.
The Protec P3000,
RC version
The Protec P3000, RC version is the standard Protec P3000 but with an external
display unit. It requires the SL3000 sniffer line.
The Protec P3000XL
The Protec P3000XL is the
HIGH
FLOW
version of the Protec P3000. It is able to detect
leaks at a much higher distance from the potential leak at reduced sensitivity but can
be switched back to normal flow at increased sensitivity. In order to use both flow
modes (high and low) it requires the use of the SL3000XL sniffer line. It may also be
operated with the normal SL3000 sniffer line, in this case the high flow mode is
disabled, however.
The Protec P3000XL,
RC version
The Protec P3000XL, RC version is identical to the Protec P3000XL, but with an
external display unit. It requires the SL3000XL sniffer line.
1.1.3
Technical Data
Physical Data
Lowest detectable leak rate
For Protec P3000
1 x 10
-7
mbar l/s
For Protec P3000XL in
LOW
FLOW
mode
1 x 10
-7
mbar l/s
For Protec P3000XL in
HIGH
FLOW
mode
1 x 10
-6
mbar l/s
Measurement range
For Protec P3000
5 decades
For Protec P3000XL in
HIGH
FLOW
mode
4 decades
Helium sensor
Wise Technology
TM
Sensor
Sensor response time
450 ms