Support by inficon, 2 support by inficon – INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 13

General Information
1-3
k
in
a
26
e Ch
ap
te
r
1
.fm
Tech
ni
ca
l Ha
nd
bo
ook
(1
20
9)
Electrical Data
Other data
1.2
Support by INFICON
INFICON Service
If equipment is returned to INFICON or an authorised INFICON representative
indicate whether the equipment is free of substances damaging to health or whether
it is contaminated. If it is contaminated also indicate the nature of the hazard.
INFICON must return any equipment without a Declaration of Contamination to the
sender’s address. You will find an appropriate form below.
Gas flow through the capillary
For Protec P3000
260 - 360 sccm*
For Protec P3000XL in
HIGH
FLOW
mode
2660 - 3500 sccm*
Time until ready for operation
approx. 3 min
* Measured at 1 atm (1013 mbar) at sea altitude. Actual flow may vary with increased
altitude and low atmospheric pressure.
Mains voltages and mains frequencies
90 - 127 V, 50 / 60 Hz
(fixed)
115 - 140 V, 60 Hz
187 - 265 V, 50 / 60 Hz
Power consumption
200 VA
Type of protection
IP 20
Overvoltage category
II
Mains cord
2.5 m
Noise level
< 54 dBA
Dimensions (w x h x d) in mm
610 x 370 x 265
Weight
27 kg
Permissible ambient temperature
(during operation)
10 °C to 45 °C
Permissible storage temperature
-40 °C to 60 °C
Relative humidity
max. 80% for temperature up to +31°C,
decreasing linearly to 50% at +40°C
Contamination level
II
(according to IEC 61010 / Part 1:
“normally only non-conductive pollution
may occur. Occasionally, however, a
temporary conductivity caused by
condensation can be tolerated.”)
Max. altitude above sea level
2000m