Flow limits

HIGH

FLOW

In this submenu the

UPPER

FLOW

LIMIT

and

LOWER

FLOW

LIMIT

for the

HIGH

FLOW

mode

may be set-up. The

LOWER

FLOW

LIMIT

can be set to values between 1500 and 2400

sccm. Default value: 1800

The

UPPER

FLOW

LIMIT

can be varied between 3000 and 3990 sccm.

Default value: 3950

Notice

To be able to measure in H

IGH

F

LOW

and L

OW

F

LOW

mode, the SL3000XL

is equipped with two capillaries (~300sccm and ~2700 sccm). In L

OW

F

LOW

mode only the smaller capillary is used, in H

IGH

F

LOW

mode both are used,

but only the gas from the smaller one is directed to the Wise Technology
sensor.

To monitor the correct flow through the smaller capillary in

HIGH

FLOW

mode

(if blocked, no leaks are detected any more), the flow limits in H

IGH

F

LOW

mode are converted into corresponding

PRESSURE

LIMITS

at the Wise

Technology sensor. If these

PRESSURE

LIMITS

are exceeded or undercut,

this is an indication of the smaller capillary’s flow being out of the normal
range and a warning message (W41 or W42) is issued.

Additional low flow error (only available in ADVANCED mode)

Notice

This submenu is only available when in

ADVANCED

mode and if a SL3000XL

sniffer line is connected to a Protec P3000XL main unit.

In this submenu an

ADDITIONAL

LOW

FLOW

ERROR

may be

ENABLED

or

DISABLED

. If

ENABLED

an additional flow error (E55 / E56) will be issued in addition to the standard

low flow warnings. Only if the

ADDITIONAL

LOW

FLOW

ERROR

is

ENABLED

, the button

LIMITS

LOW

FLOW

ERROR

will be available.

If a

LOW

FLOW

ERROR

occurs, the Protec P3000XL will stop measuring. Only when the

cause of the low flow problem has been resolved and the flow is above the

LIMITS

LOW

FLOW

ERROR

again, the Protec P3000XL will go back into measurement mode.

The

LIMITS

for the

LOW

FLOW

ERROR

may be set -up in the

LIMITS

LOW

FLOW

ERROR

submenu (see below).

Default setting: disabled

Fig. 4-12 Variation of flow limits