Wait time

The

WAIT

TIME

between two leak checks allowed to move the sniffer to the next

potentially leaking location can be varied between 0.1 s and 25.0s in steps of 0.1s .
Please use the

UP

and

DOWN

buttons for editing this parameter.

Default value: 3.0 s

Global trigger

In the

GLOBAL

TRIGGER

submenu the maximum allowable leak rate for the complete

unit under test can be edited. The same ranges as for a single trigger value are
allowed. The unit of measurement for the global trigger will be the same as selected
for this gas type in the edit gas submenue.

Default value: 2 x 10

-3

mbar l/s (or the equivalent in any other unit of measurement)

4.4.5

Miscellaneous Settings

Language

Different languages can be chosen in which the software can be operated. The
current selection includes English, German, Spanisch, Italian, French, Portugese,
Japanese (Katakana) and Chinese (Mandarine).

Default value: English

Notice

If the Protec P3000 has been set to a language that you do not understand,
turn off the power and press the second button from the top of the display
on both sides simultaneously during run-up. This will reset the Protec
P3000 to English as the menu language. This setting is not saved
automatically. Please go to the

LANGUAGE

submenu afterwards and select

the appropriate language.

Date & Time

In this submenu the internal clock of the Protec P3000 can be set. On the first page
the date needs to be entered in the format DD.MM.YYYY. The lower right button (

→)

opens the second page on which the time in the format HH.MM is entered.

Sniffer light

The intensity of the sniffer light can be varied to the desired level. The intensity can
be selected between 1 and 6. Also the light can be switched on and off via the
D

ISABLE

/ E

NABLE

buttons. Any new setting must be confirmed with the OK button.

Default value: on, level 4.

Pressure unit

The pressure unit of all displayed values of the Protec P3000 can be set to Pa, Torr,
atm or mbar.

Default value: mbar