INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 24
1-14
General Information
k
in
a
26
e Ch
ap
te
r
1
.fm
Tech
ni
ca
l Ha
nd
bo
ook
(1
20
9)
PRO-Check Reference
leak (optional)
Please insert the PRO-Check reference leak into the opening in the housing of the
main unit. Make sure that the Sub-D plug is properly connected with the PRO-Check
leak.
Notice
When properly inserted, the PRO-Check reference leak will still protrude by
approx. 10 mm.
On first usage of your PRO-Check you need to initialize the use of this reference leak
in the Protec P3000 software.
Please perform the following steps:
1
Insert the PRO-Check into the appropriate opening of the Protec P3000
2
In the software menu go to
HISTORY
&
MAINTENANCE
/
REPLACE
PRO
-
CHECK
.
Notice
This menu item is only available when the Protec P3000 is set to
ADVANCED
MODE
(see Section
, Keyword: User Mode)
3
On the certificate, which is delivered with the PRO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open
submenu and the 12-digit-code in the second line and press OK.
Notice
The PRO-Check reference leak must be installed in the Protec P3000
when pressing
OK
.
Notice
PRO-Check Warntime Expire Date (See Section 7.7.5).
For RC versions only
The RC version has no built-in display unit but a connectors plate is mounted instead.
Please connect the external display unit with the 5m connecting table (Cat.-no. 551-
002)
Fig. 1-5 Initializing the PRO-Check reference leak