beautypg.com

INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 50

background image

3-16

Operation of the Protec P3000

k

in

a

26

e Ch

ap

te

r

3

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

If the value measured for the PRO-Check is out of range, the message "recalibration
required!" will be displayed in inverted colors.

For returning to the measuring mode, please press the right probe button or press

OK

” on the main display.

For Protec P3000XL only

For the Protec P3000XL both calibration factors for

HIGH

FLOW

and

LOW

FLOW

mode

will be verified (see fig. 3-16). However, only the information for the selected flow
mode will be relevant for passing or failing the verification. The results for the non-
selected mode will be displayed for reference at the bottom of the results screen.

Fig. 3-16 Results of proof function for Protec P3000

Fig. 3-17 Results of proof function for Protec P3000XL