Despatch LCC/LCD2-14-3 Non Pass Through Series User Manual

Page 72

background image

63

Control

Menu Item

Display

Default

Range

Setting

Cycle Time

Control Cycle Time

1

1 to 60 seconds

High limit setpoint

Control Hi-Lim SP

Max HiLimSP

MinHiLimSP - MaxHiLimSP*

High limit band

Control Hi-Lim Band

Off

Off, 3°C to 11°C (5°F to 20°F)

Power fail recovery

Control PwrFRec

Stop

Stop, Restart, Hold, Resume

Recovery time limit

Control PwrFTime

00m00s

00m00s to 99m59s

Powerup start enable

Control EPwrStrt

Dis

Dis, En

Powerup Start Mode

Control StrtMode

Off

Off, Manual, Timer, Pro-1 to Pro-8

Hysteresis

Control Hyst

3°C

1°C to 56°C (1°F to 100°F)

Process out low

Control RetOutLo

80°C

-73°C to 760°C (-100°F to 1400°F)

Process out high

Control RetOutHi

400°C

-73°C to 760°C (-100°F to 1400°F)

Time scale

Control TimeScale

hh:mm

hh:mm or mm:ss

Key press beep

Control KeyBeep

On

On or Off

End of cycle beep

Control EOCBeep

Off

On or Off

Alarm beep

Control AlarmBeep

Off

On or Off


*see Enable Page

Communication (optional)

Menu Item

Display

Default Range

Setting

Address

Communication CommAddr

1

1 to 247

Mode

Communication CommMode

OFF

OFF, Modbus

Baud rate

Communication Baud Rate

19.2K

2400, 4800, 9600,
19.2K, 38.4K

Parity

Communication Parity

None

None, Odd, Even