LCC/LCD2-14-3 POWER & EXHAUST CONNECTIONS
WARNING:
Never allow drain to be plugged as a hot oven will generate a small amount of steam
when the water is first turned on. STEAM BURNS!
CAUTION:
Design the drain system to prevent operator injury from high temperature or pressure
buildup. Piping must be able to withstand short periods of up to 650 °F (343 °C)
temperatures. Drain lines should be insulated and/or warning labels installed that a
hazard exists.
Conduit fitting for
customer power line.
Conduit runs up to
front control panel.
Wires terminate at
disconnect switch.
Cabinet cooling fan, air
blowing out. DO NOT
COVER! Required for
lower compartment
ventilation.
4.00” (10.16CM) diameter
exhaust opening. Exhaust
temperatures may reach 260ºC
on LCC versions, and 350ºC on
LCD versions. Make sure
exhaust piping materials can
withstand these temperatures.