PID Page


The PID Page contains parameters which control the response to the setpoint and
process variable input. To access the PID Page, enter the Setup Mode. Press the Page
key until PID is displayed. Press the Menu key. Each parameter can be changed by
pressing the Menu key until the desired parameter is displayed, and then pressing the

keys to change the value.

Menu Item

Display

Description

Range

Display units

PID Temp Unit

Set display units to °C or °F

°C or °F

Proportional
band

PID Prop Band

Set proportional band for tuning

1 to 56°C (1 to 100°F)

Integral reset

PID
Reset/Rep/Min

Set integral reset for tuning

0.0 to 100 seconds/repeat

Derivative
Rate

PID Rate In Sec

Set derivative rate for tuning

0.0 to 500 seconds

AutoTune

PID AutoTune

Enable auto tuning function

Disable, Enable



The AutoTune parameter disables or enables the AutoTune function. To utilize
AutoTuning: