PR-8V High Performance Thickness Gauge

71

Note: In the photos above, the MEM format is set to GRID. However, if the

ENGINE format was selected, the COL would change to CYL#, and the ROW

would be displayed as CLK#.

1) Press the MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left until the COL or ROW cell is highlighted.

2) Press the ENTER key to display the Grid View Box.

3) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the stored

readings and corresponding display view. Notice as the cursor is moved to

a different cell, the display will be updated with the display view saved with

the reading.

4) The user may opt to clear a specific reading and save a new one at any

time. Press the CLR key in the appropriate cell location to clear the reading,

take a new measurement, and press the OK key to save the new reading.

5) Abort the Grid View Box by pressing the MEAS key at any time.

10.6 Deleting Grids (Files)

Deleting One or All Files from Memory

Deleting One File