beautypg.com

Dakota Ultrasonics PR-8V User Manual

Page 42

background image

Dakota Ultrasonics

38

1) Using a set of calipers, physically measure two known thickness locations

on the sample of material. Note: The two locations should represent the

desired thickness range being measured. If the range is from .250” - 1.000”,

then the two known locations should represent this range, or be slightly

outside or inside of this range. A typical example would be to find two

locations at .236” and the other at 1.047”. These thickness points will satisfy

the required thickness range noted above.

2) Apply a drop of couplant on the transducer. Place the transducer in steady

contact on thickest known location first. Be sure that the reading is stable

and the repeatability indicator, located in the top left corner of the display, is

fully lit and stable. Press the MENU key once to activate the menu items tab.

Press the MENU key multiple times to tab right, and the ESC key multiple

times to tab left, until the CAL menu is highlighted and displaying the

submenu items.

3) Use the UP and DOWN arrow keys to scroll through the sub menu items

until ONE POINT is highlighted.

4) Press the ENTER key to display the Digits Edit Box.

5) Press the UP and DOWN arrow keys to scroll the highlighted value.

6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity for the ONE POINT calibration

and return to the CAL menu screen, or press ESC to cancel entering the

velocity.

9) Apply a drop of couplant on the transducer and place the transducer in

steady contact with the sample or actual test material on the thinnest, or