beautypg.com

Dakota Ultrasonics PR-8V User Manual

Page 24

background image

Dakota Ultrasonics

20

The top & bottom end panels are where all connections are made to the PR-8V. The
diagram above shows the layout and description of the connectors:

Transducer Connectors

Refer to Diagram: The transducer connectors, and battery cover/probe zero disk are
located on the PR-8V’s top end cap. The transducer connectors are of type Lemo
“00”. Note: There is no polarity associated with connecting the transducer to the PR-
8V
.

Probe Zero Disk & Battery Cover

Refer to Diagram: The Battery cover is the large round disk shown in the diagram.
Note: This same disk is also used as a probe zero disk. Simply remove the cover
when replacing the batteries (3 AA cells). When performing a probe zero function,
simply place the transducer on disk making firm contact. Important: Be sure to
follow the polarity labels located on the back label of the PR-8V. Note:
Rechargeable batteries can be used, however they must be recharged outside of the
unit in a stand alone battery charger.

RS-232 Connector

Refer to Diagram: The RS-232 connector, located on the bottom end cap, is a 2 pin
female Lemo connector. It is designed to connect directly from the PR-8V to a
standard AT serial port on a PC. The cable supplied with the PR-8V is a Lemo to 9
pin serial cable. Note: This connector is also used to upgrade the PR-8V with the
latest version of firmware.