2 alarm mode, Toggle alarm status (on/off) – Dakota Ultrasonics PR-8V User Manual
Page 61
PR-8V High Performance Thickness Gauge
57
9.2 Alarm Mode
The Alarm Mode feature of the PR-8V provides the user with a method of setting
tolerances, or limits, for a particular application requirement. This feature may be
used for a variety of applications to verify the material is within the manufacturer
specifications. There are two limits, or alarm values, that can be setup in the PR-8V
– ALARM LO LIMIT and ALARM HI LIMIT . However, the user may choose to
activate and utilize only one of the limit values, depending on their specific
application requirements. The procedures to use the ALARM MODE feature are
outlined below:
UTIL MENU – ALARM MODE
Toggle Alarm Status (on/off)