beautypg.com

5 user control and mode of operation, Buser control and mode of operation – Eppendorf PatchMan NP 2 User Manual

Page 28

background image

73

5.2.4 Capillary exchange

Before swivelling it out, make sure the capillary cannot be impeded by any obstacles!
Wear goggles to protect against possible glass splinters! Swivel out module unit on
Y/Z module!

Swivel module unit for capillary exchange out of working radius. The capillary swivels out
forwards and upwards with the module unit.

Before changing the capillary, bring the capillary to a position over the petri dish either
manually or by way of the Home function.

Grip module unit at rear motor unit (Y/Z module) and swivel out.

Do not use the X module!

Exchange capillaries and swivel module unit back into place.

Bring the new capillary back into the focal plane.

Fig. 5: Module unit in working position and swivelled out
(here without capillary holder/preamplifier)

0

40°

40°

20°

20°

5 User control and mode of operation

BUser control and mode of operation

5

05_Bedien_e.fm Seite 73 Dienstag, 16. August 2005 3:18 15