Calibration examples – WIKA Pascal ET User Manual
Page 80
6 Commissioning, operation
80 WIKA Operating Instruction, Pascal ET and Pascal ET/IS
GB
6.9
Calibration Examples
Temperature calibration with the comparison method
DESCRIPTION
The user has to calibrate a temperature element (RTD or TC) in the field.
Accordingly, customer needs:
■
to measure the signal coming from the RTD or TC under test (DUT)
■
to measure the reference signal coming from the reference RTD or TC
■
to have a temperature generator
UP TO NOW
User has to carry in the field:
1. a standard calibrator to measure the TC or
RTD signal coming from the DUT
2. an external indicator to measure the
reference signal coming from the reference
TC or RTD
3. a dry block to generate the temperature
FROM NOW
User can use :
1. the PASCAL ET II ( 2 Input) to:
■
measure the TC or RTD signal coming
from the DUT
■
measure the reference signal coming
from the reference TC or RTD
2. a dry block to generate the temperature
ADVANTAGES
1) customer can buy only ONE CALIBRATOR, d
oesn’t need to buy two devices
– COST SAVINGS
2) customer has to carry in the field only ONE CALIBRATOR and no more two
devices
– EASIER JOB
3) customer can run the calibration directly on the calibrator, storing the
calibration data
– FASTER CALIBRATION – TIME & COST SAVINGS