6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET and Pascal ET/IS 19
GB
6.1.2.3 Pressure module
It is possible to install two pressure modules, each one can support one external transducer.
Accordingly 2 external sensors can be connected at the same time.
Transducers selection with range and their resolutions are in the following table.
External transducers *
Range
Precision (% FS)
Accuracy (% FS)
Typical Resolution
Gauge
-
60…+60 mbar
0,1
0,15
0,01 mbar
-
500…+500 mbar
0,015
0,025
0,01 mbar
-
900…+1500 bar
0,015
0,025
0;01 mbar
0…7 bar
0,015
0,025
0,1 mbar
0…21 bar
0,015
0,025
0;1 mbar
0…50 bar
0,015
0,025
1 mbar
0…100 bar
0,015
0,025
1 mbar
0…200 bar
0,015
0,025
10 mbar
0…400 bar
0,015
0,025
100 mbar
0…700 bar
0,025
0,05
100 mbar
0…1000 bar
0,025
0,05
100 mbar
Absolute
0…1500 mbar abs.
0,015
0,025
0,01 mbar
0…2500 mbar abs
0,015
0,025
0,01 mbar
0…5 bar abs.
0,015
0,025
0,1 mbar
0…7 bar abs.
0,015
0,025
0,1 mbar
0…21 bar abs.
0,015
0,025
0,1 mbar
0…81 bar abs.
0,015
0,025
1 mbar
0…100 bar abs.
0,015
0,025
1 mbar
* Other ranges available on request
6.1.2.4 HART module
The HART module allows communication with HART instruments (usually transmitters) permitting
to acquire digital measurements, instrument data and modifying settings. It can be inserted in the
front panel in place of OUT A or OUT B module.
It is a fully galvanically isolated board (like IN and OUT modules) from internal electronics of the
Pascal.
The module allow the 24V to supply directly the output loop for the transmitter.
Further, it can supply a 250Ω load resistance electronically switchable necessary to HART
communication, permitting to avoid the use of an external one.
The following figure shows the pin connections in HART module:
Figure 3
– HART module