6 Commissioning, operation

52 WIKA Operating Instruction, Pascal ET and Pascal ET/IS

GB

Calculation is possible only when channel 1 is assigned for mV measurement (bridge signal) and
channel 2 is assigned for V generation, range 0÷20 V (bridge supply). Moreover channel 1 should
not be set in error mode.
In this case, below what will be displayed:

Figure 60

– Cell Load calculation


Pascal ET or Pascal ET/IS allows to manage the sensitivity of the load cell from 0 mV/V up to
9.99999 mV/V.
Moreover, on channel 4, the SCALING function can be assigned through the menu CHANNEL in
order to display an engineering unit more appropriate (for example Kg or a pressure unit).

Figure 61

– Cell Load calculation with Scaling function active