Environmental parameters module (optional), Transmitter hart slave – WIKA Pascal ET User Manual
Page 22
![background image](/manuals/465911/22/background.png)
6 Commissioning, operation
22 WIKA Operating Instruction, Pascal ET and Pascal ET/IS
GB
Figure 8 - Case 5: 24V from m
odule, 250Ω from module, mA reading by IN module
6.1.2.5 Environmental parameters module (optional)
This module allows to measure: temperature, relative humidity and atmospheric pressure. These
parameters represent the most important factors that affect the calibration results. This module is
connected to Pascal ET or Pascal ET/IS with a multiple pin connector present on the right side
panel. This module is plug & play and the relevant information about the environmental parameters
are displayed on the status bar at the top of the screen.
This information is automatically reported on the calibration report.
Parameter
Range
Accuracy
Uncertainty
Max Resolution
Temperature
(-10 ÷ 50) °C
1,5 °C
1,8 °C
0,1 °C
Barometric Pressure (650 ÷ 1150) mbar
4 % fs
5 % fs
1 mbar
Relative Humidity
10 % ÷ 90 %
4 %
5 %
1 %
Transmitter
HART slave