6 Commissioning, operation

34 WIKA Operating Instruction, Pascal ET and Pascal ET/IS

GB

Once RTD USER is selected, it is possible to set the parameter of the RTD used.


A

=

X * 10

-3

°C

-1

B

=

X * 10

ˉ7

°C

-2

C

=

X * 10

-12

°C

-3

Ro

=

Resistance @ t = 0,01 °C

X

=

Value


Next step is to define the measurement range. The range selection is necessary during the
calibration procedure to define the relationship between the REF and DUT channels.

Figure 20

– Range setup


Now it is possible to conclude the procedure by pressing the YES button or come back to the
previous step pressing NO.

Figure 21

– Channel setup confirmation

Once the YES key is pressed, the calibrator returns to the initial page where the user can proceed
with the measurement/simulation, or assign another function to a different software channel
following the same procedure.