beautypg.com

WIKA Pascal ET User Manual

Page 76

background image

6 Commissioning, operation

76 WIKA Operating Instruction, Pascal ET and Pascal ET/IS

GB

Pressing POINTS, the list of the calibration points will be shown:

Figure 93

– Display test points Report


Pressing RESULT, the calibration results are shown:

Figure 94

– Display Report results


Press the

“scrolling menu”, menu with “As found”, “As left”, “Up”, “Down”, will appear to enable to

see the calibration results related to the different calibration stages.
“As found”, (calibration points before any adjustment on the DUT) “As left” (calibration points after
the adjustment on the DUT has been carried out)

,“Up” (calibration points going up), “Down”

(calibration points going down).
Pressing GRAPHIC, the graphic of the calibration will be displayed, see below:

Figure 95

– Display Report in graphic mode

Scrolling menu