beautypg.com

Myron L D-6 User Manual

Page 41

background image

38

COrrECTIVE ACTION
• Check battery connections or replace battery.

(See Battery Replacement, pg. 32.)
• Recalibrate pH function. (See pH Calibration (D-6), pg. 20.)
• Thoroughly rinse sensor well.

• Recalibrate using fresh buffers.

(See pH Calibration (D-6), pg. 20.)
• Replace pH/ORP Sensor.

(See pH/ORP Sensor Replacement (D-6), pg. 33.)

• Clean and rejuvenate sensor. (See pH/ORP, pg. 35.)

• Recalibrate. (See pH Calibration (D-6), pg. 20.)

• If no improvement, replace pH/ORP sensor.

(See pH/ORP Sensor Replacement (D-6), pg. 33.)
• Clean and rejuvenate sensor. (See pH/ORP, pg. 35.)

• Recalibrate. (See pH Calibration (D-6), pg. 20.)

• If no improvement, replace pH/ORP sensor.

(See pH/ORP Sensor Replacement (D-6), pg. 33.)

• Clean cell cup and electrodes. (See Cleaning Sensors, pg. 34.)

• Minimize test sample exposure to air by using a flowing sample.

Let sample flow continuously into conductivity cell with no aeration.
• Clean cell cup and electrodes. (See Cleaning Sensors, pg. 34.)

• Minimize test sample exposure to air by taking a flowing sample.

Let sample flow continuously into conductivity cell with no aeration.

• Clean cell cup and electrodes. (See Cleaning Sensors, pg. 34.)

• Rinse cell cup more thoroughly before measurement.

• Ensure pH cap is snugly in place.
• Minimize test sample exposure to air by taking a flowing sample.

Let sample flow continuously into conductivity cell with no aeration.