Myron L D-6 User Manual

Page 28

background image

2.

Rinse and fill pH/ORP sensor well with pH Sensor Storage

Solution immediately after each use.

NEVEr use distilled water or tap water as a substitute for pH

Sensor Storage Solution. This will deplete the sensor quickly,

significantly decreasing its lifespan.

• If pH Sensor Storage Solution is unavailable, use:

• An almost saturated KCl solution (preferred)
• A saturated solution of NaCl (table salt) and tap water

(next best alternative)

• pH 4 buffer (least desirable alternative).

• Replace with Myron L pH Sensor Storage Solution as soon

as possible.

• Make sure the pH/ORP sensor cap is reinstalled securely

over the pH/ORP sensor well.

ORP calibration solutions are caustic, and ±5% is considered very

accurate. By using the pH zero setting (0 mV = 7 pH) for ORP and

precision electronics for detection, the D-6 delivers better accuracy without

calibration than a simpler instrument could using calibration solutions.

XII. MEMOrY
This feature allows up to 100 readings with their temperatures to be stored

simultaneously for later recall. At the same time, the TIME and DATE are

also recorded. Memory download to a computer requires the bluDock™

accessory package. Order MODEL # BLUDOCK.

A. Memory Storage

1.

While displaying a

measurement, press

to record the

displayed value.

2.

“MEMORY” will appear

and the temperature

display will be momentarily replaced by a number (1-100)

showing the position of the record. Figure 8 shows a reading

of 1806 µS stored in memory record #4.

B. Memory Recall

1.

Press any measurement key.

2.

Press

, “MEMORY” will appear, and the display will show

the last record stored.

Figure 8

°C

KCl

COND

MEMORY

25