Myron L D-6 User Manual
Page 29

3.
Press
or
to scroll to the record location desired
(the temperature display alternates between temperature
recorded and location number).
4.
Press
to display time and date stamp.
5.
Press any measurement key to leave memory recall or allow to
automatically turn off.
C. Clearing a Record/Memory Clear
After recalling a certain record location, press and hold
to
clear that memory. This space will be the place for the next memory
record, unless you scroll to another empty position before ending the
recall sequence. The next measurement stored will go into the next
highest available memory location.
Example: You have locations 1-7 filled and wish to clear the conductivity
reading stored in record location “3” and replace it with a pH reading.
1.
Press
and scroll to location “3”.
2.
Press and HOLD
to clear old record “3”.
3.
Fill pH/ORP sensor well with sample.
4.
Press
to measure sample and press
to store
reading in location “3”.
5.
The next memory stored will go into location “8”.
6.
To clear all records: After pressing
, scroll down.
“CL
r
ALL” will be displayed (see Figure 9).
7.
Press
. All records will be cleared.
XIII.
TIME and DATE
The Time and Date may easily be
changed.
A. Setting TIME
Time is always displayed in 24 hour
time.
Figure 9
MEMORY
26