B. Maintenance of the pH/ORP Sensor

The sensor well must be kept wet with a solution. Before replacing

the rubber cap, rinse and fill the sensor well with Myron L pH Sensor

Storage Solution. If unavailable, you can use an almost saturated KCl

solution, a saturated solution of table salt and tap water or pH 4 buffer

(ref. pH and ORP Practices to Maintain Calibration, pg. 24). NEVEr

uSE DISTILLED WATEr.

VII.

SPECIFIC rECOMMENDED MEASurING PrOCEDurES

If the proper solution is not selected (KCl, NaCl, or 442), see Solution

Selection, pg. 15.

CAuTION:

• For Cond/Res/TDS measurements:

• The pH/ORP sensor well cap MuST be sealed.

• Any crystals or other residue around the cap MuST be rinsed

away to prevent contamination of the conductivity sample.

• The conductivity cell MuST be rinsed and filled at least 3 times

with sample before each measurement is taken.

• For pH/ORP measurements:

• Both the pH/ORP sensor well and the conductivity cell MuST

be rinsed and filled at least 3 times with sample before each

measurement is taken.

• After sampling high concentration solutions or temperature

extremes, more rinsing may be required.

A. Measuring Conductivity & Total Dissolved Solids (TDS)

WArNING:

When using a Digital Dialysate Meter™ to check the

conductivity of dialysate, the solution conversion factor MuST be

set to the NaCl conversion factor.

• If the proper solution conversion factor is not selected, see Procedure

to Select a Solution, pg. 15, for instructions on setting it.

1.

Rinse cell cup 3 times with sample to be measured. (This

conditions the temperature compensation network and prepares

the cell.)

2.

Refill cell cup with sample.

3.

Press

or

.

4.

Take reading. A display of [- - - -] indicates an overrange

condition.