Myron L D-6 User Manual
Page 26

XI.
CALIBrATION INTErVALS
The Dialysate Meter is designed to not require frequent recalibration. The
most common sources of error were eliminated in the design, and there
are no mechanical adjustments. Still, to ensure specified accuracy, any
instrument must be checked against chemical standards occasionally.
A. Suggested Calibration Intervals
PRIMARY
USAGE
FUNCTION TYPE OF CHECK
RECOMMENDED
FREQUENCY
Checking
Dialysate
Conductivity Check against
standard conductivity
solutions
At the beginning
and end of each
workday
Check cell cup
cleanliness
At the beginning
and end of each
workday
pH
Check against
standard buffer
solutions
At the beginning
and end of each
workday
Check sensor well
cleanliness
At the beginning
and end of each
workday
Temperature No calibration required
—————
Checking
Water
Quality
Conductivity Check against standard
conductivity solutions
Monthly
Check cell cup
cleanliness
Monthly
Resistivity
No specific check
required — calibration
based on Conductivity
calibration
—————
TDS
Check against
standard solution
Monthly
pH
Check against standard
buffer solutions
Every 2 weeks
Check sensor well
cleanliness
Every 2 weeks
ORP
Automatically calibrated
during pH calibration
—————
Temperature No calibration required
—————
23