XI.

CALIBrATION INTErVALS

The Dialysate Meter is designed to not require frequent recalibration. The

most common sources of error were eliminated in the design, and there

are no mechanical adjustments. Still, to ensure specified accuracy, any

instrument must be checked against chemical standards occasionally.

A. Suggested Calibration Intervals

PRIMARY

USAGE

FUNCTION TYPE OF CHECK

RECOMMENDED

FREQUENCY

Checking

Dialysate

Conductivity Check against

standard conductivity

solutions

At the beginning

and end of each

workday

Check cell cup

cleanliness

At the beginning

and end of each

workday

pH

Check against

standard buffer

solutions

At the beginning

and end of each

workday

Check sensor well

cleanliness

At the beginning

and end of each

workday

Temperature No calibration required

—————

Checking

Water

Quality

Conductivity Check against standard

conductivity solutions

Monthly

Check cell cup

cleanliness

Monthly

Resistivity

No specific check

required — calibration

based on Conductivity

calibration

—————

TDS

Check against

standard solution

Monthly

pH

Check against standard

buffer solutions

Every 2 weeks

Check sensor well

cleanliness

Every 2 weeks

ORP

Automatically calibrated

during pH calibration

—————

Temperature No calibration required

—————

23