beautypg.com

Myron L D-6 User Manual

Page 21

background image

Function

Solution

Recalibrate if

Dialysate Meter

Reports Below

Recalibrate if

Dialysate Meter

Reports Above

Conductivity

(NaCl)

14.00 mS

13.86 mS

14.14 mS

Conductivity

(KCl)

7000 µS

6930 µS

7070 µS

TDS

442™ Solution

(3000 PPM)

3030 ppm

2970 ppm

pH

7.0 Buffer

6.96 pH

7.04 pH

4.0 Buffer

3.96 pH

4.04 pH

10.0 Buffer

9.96 pH

10.04 pH

Once in “CAL” mode, the

key becomes an accept key. At each

point, pressing

accepts the new calibration value and steps you

to the next adjustment (or out of “CAL” mode if there are no more steps).
To bypass a calibration step, simply press

to accept the present

value as is.

b. Leaving Calibration

Calibration is complete when the “CAL” icon goes out. Pressing any

measurement key cancels changes not yet accepted and exits calibration

mode.

Leaving pH after the 2nd buffer results in the same gain being entered

in place of the 3rd buffer.

2. Calibration Limits

There are calibration limits. A nominal “FAC” value is an ideal value

stored by the factory. Attempts to calibrate too far, up or down, from

there will cause the displayed value to be replaced with “FAC”. If

you accept it (press the CAL key), you will have the original default

factory calibration for this measurement. The need to calibrate out

so far that “FAC” appears indicates a procedural problem, incorrect

standard solution, a very dirty cell cup or an aging pH/ORP sensor (ref.

Troubleshooting Chart, pg. 37-38).

WArNING:

• Factory settings do not take into account the current condition of

the sensor installed in your instrument.

• If you are unable to adjust the calibration value without causing

the “FAC” icon to appear, your conductivity or pH/ORP sensor is

damaged and must be replaced.

18