Myron L D-6 User Manual
Page 18

5.
Press any parameter key to exit ORP unit preference selection
or let the unit time out. ORP unit preference will be saved.
NOTE: Free chlorine as determined by the D-6 Dialysate Meter is a
measurement of the concentrations of Cl
2
, HOCl, and OCl- in solution. The
D-6 Dialysate Meter does NOT measure chloramines present in solution.
IX.
SOLuTION SELECTION
A. Why Solution Selection is Available
Conductivity, resistivity, and TDS require temperature correction to 25°C
values. Selection determines the temperature correction of conductivity
and calculation of TDS from compensated conductivity.
B. The 3 Solution Types
On the left side of the display is the salt solution characteristic used
to model temperature compensation of conductivity and its TDS
conversion. In hemodialysis facilities, using NaCl for Conductivity and
Resistivity and 442 (Natural Water characteristic) for TDS will meet
water testing requirements most of the time.
This assumes that Conductivity will be used to determine concentration
of dialysate solutions and that Resistivity will be used to determine
the purity of DI water given both dialysate and DI product water most
closely match the NaCl standard. This also assumes that TDS will be
used to check reverse osmosis feed and product water, which most
closely match the Natural Water characteristic.
This is how your instrument, as shipped from the factory, is set to
operate. However, the compensation and conversion solution type
can be changed to another programmed solution characteristic, when
necessary.
C. Calibration of Each Solution Type
A Digital Dialysate Meter™ stores and uses separate calibration values
for each of the three solution types.
If you’re using a Digital Dialysate Meter™ for BOTH dialysate
measurements AND product water quality measurements, ALWAYS
calibrate the conductivity, resistivity and TDS functions of your Digital
Dialysate Meter™ FOr EACH TYPE.
• Calibrating a function for one solution type DOES NOT affect
the calibration of the same function for other solution types.
• For example, calibrating a measurement function, such as
conductivity, for NaCl DOES NOT calibrate conductivity for KCl
or 442™.
D. Procedure to Select a Solution
WArNING:
When using a Digital Dialysate Meter™ to check the
conductivity of dialysate, the solution type MuST be set to use the
15